2640A/2645A Combined Specifications

A

A-5

Table A-3. 2640A/2645A DIGITAL I/O Specification (cont)

Specification

Characteristic

Output Voltage - TTL Logical Zero

0.8V maximum for an I out of -1.0 mA (1 LSTTL

load)

Output Voltage - TTL Logical One

3.8V minimum for an I out of 0.05 mA (1 LSTTL
load)

Output Voltage - Non-TTL Load Zero

1.8V maximum for an Iout of -20 mA

Output Voltage - Non-TTL Load One

3.25V maximum for an Iout of -50 mA

Trigger In

A-7.

Table A-4 provides a summary of the Trigger In specifications. The Trigger In
input is located on the ALARM/TRIGGER I/O connector, terminals TI and GND.

Table A-4. 2640A/2645A Trigger In (TI) Specification

Specification

Characteristic

Logical High - Trigger not set

Minimum: 2.0V

Maximum: 7.0V

Logical Low - Trigger set

Minimum: -0.6V
Maximum: +0.8V

Compatibility

TTL or Contact Closure

Isolation

None (dc coupled)

Minimum Pulse Width

5 µs

Maximum Frequency

Nominal 400 Hz

Repeatability

3 ms