Trigger out, Master alarm, Trigger out a-8 – Fluke NetDAQ 2645A User Manual

Page 216: Master alarm a-9

Advertising
background image

2640A/2645A NetDAQ

Users Manual

A-6

Trigger Out

A-8.

Table A-5 provides a summary of the Trigger Out specifications. The Trigger Out
output is located on the ALARM/TRIGGER I/O connector, terminals TO and
GND.

Table A-5. 2640A/2645A Trigger Out (TO) Specification

Specification

Characteristic

TTL Logical Zero - Trigger Out Set

0.8V maximum for an Iout of -1.0 mA (1 LSTTL
load)

TTL Logical One - Trigger Out Not Set

3.8V minimum for an Iout of 0.05 mA (1 LSTTL
load)

Non-TTL Logical Zero - Trigger Out Set

1.8V maximum for an Iout of -20 mA

Non-TTL Logical One - Trigger Out Not Set

3.25V maximum for an Iout of -50 mA

Pulse Duration (Logic Low)

125 µs

Isolation

None

Master Alarm

A-9.

Table A-6 provides a summary of the Master Alarm specifications. The Master
Alarm output is located on the ALARM/TRIGGER I/O connector, terminals MA
and GND.

Table A-6. 2640A/2645A Master Alarm (MA) Specification

Specification

Characteristic

TTL Logical Zero - Master Alarm Set

0.8V maximum for an Iout of -1.0 mA (1 LSTTL
load)

TTL Logical One - Master Alarm Not Set

3.8V minimum for an Iout of 0.05 mA (1 LSTTL
load)

Non-TTL Logical Zero - Master Alarm Set

1.8V maximum for an Iout of -20 mA

Non-TTL Logical One - Master Alarm Not Set

3.25V maximum for an Iout of -50 mA

Isolation

None