2640A Specifications

A

A-7

2640A/2645A Totalizer

A-10.

Table A-7 provides a summary of the Totalizer specifications. The Totalizer input
is located on the DIGITAL I/O connector, terminals

and GND.

Table A-7. 2640A/2645A Totalizer Specification

Specification

Characteristic

Maximum Input Voltage

30V

Minimum Input Voltage

-4V

Minimum Peak Voltage

2V

Isolation

None (dc coupled)

Threshold

1.4V

Hysteresis

500 mV

Input Debouncing

None or 1.75 ms (selectable)

Maximum Transition Rate

5 kHz (Debounce disabled)
500 Hz (Debounce enabled)

Maximum Count

4,294,967,295

2640A/2645A Real-Time Clock and Calendar

A-11.

Table A-8 provides a summary of the battery powered real-time clock and
calendar.

Table A-8. 2640A/2645A Real-Time Clock and Calendar

Specification

Characteristic

Accuracy

1 minute per month for 0

°

C to 50

°

C range

Battery Life

>15 unpowered instrument years for -20°C to
28

°

C (32

°

F to 82.4

°

F).

>6 unpowered instrument years for -20

°

C to

50

°

C (32

°

F to 122

°

F).

>4 unpowered instrument years for 50

°

C to

70

°

C (122

°

F to 158

°

F).

2640A Specifications

A-12.

This section includes specifications specific to the 2640A instrument by
measurement function.