2640a dc voltage measurement specifications – Fluke NetDAQ 2645A User Manual

Page 218

Advertising
background image

2640A/2645A NetDAQ

Users Manual

A-8

2640A DC Voltage Measurement Specifications

A-13.

Tables A-9 to A-11 provide 2640A specifications for the dc voltage measurement
function.

Table A-9. 2640A DC Voltage Measurement General Specifications

Specification

Characteristic

Input Impedance

100 M

in parallel with 150 pF maximum for

ranges

3V

10 M

in parallel with 100 pF maximum for ranges

>3V

Normal Mode Rejection

50dB minimum at 50 Hz/60 Hz

±

0.1%, Slow Rate

Common Mode Rejection

120dB minimum at dc, 50 Hz/60 Hz

±

0.1%, 1 k

imbalance, Slow Rate

80dB minimum at dc, 50 Hz/60 Hz

±

0.1%, 1 k

imbalance, Medium and Fast Rates

Channel-to-Channel Crosstalk

120dB minimum Slow Rate (e.g., 30V dc on
channel 1 may cause a 30µV error on channel 2)

100dB minimum Medium and Fast Rates (e.g., 1V
dc on channel 1 may cause a 10 µV error on
channel 2)

Temperature Coefficient

Add 1/10th the 90-day specification per

°

C above

28

°

C or below 18

°

C. (Generally, only the %input

portion is affected.)

Accuracy at -20

°

C

Multiply the -10

°

C to + 60

°

C accuracy specification

by 2. After 1 hour warm-up. For accuracy between
-10

°

C and -20

°

C, interpolate linearly.

Maximum Input Voltage

150V (300V for channels 1 and 11) to any input
terminal.

Table A-10. 2640A DC Voltage Range and Resolution Specifications

Resolution

Range

Slow

Fast

90 mV

.3 µV

1 µV

300 mV

1 µV

3 µV

3V

10 µV

30 µV

30V

100 µV

300 µV

150V/300V

1 mV

3 mV

Note 300V range applies to channels 1 and 11 only.