2640A Specifications

A

A-9

Table A-11. 2640A DC Voltage Accuracy Specifications

Accuracy, 3

σ

±

(% input + V)

18

°

C to 28

°

C

-10

°

C to 60

°

C

Range

90 Day

1 Year

1 Year

Slow

Fast

Slow

Fast

Slow

Fast

90 mV

.01%+7 µV

.01%+17 µV

.013%+8 µV

.013%+18 µV

.042%+18.2µV .042%+44.2µV

300 mV

.01%+15 µV

.01%+30 µV

.013%+17 µV

.013%+35 µV

.042%+39 µV

.042%+78 µV

3V

.01%+.1 mV

.01%+.2 mV

.013%+.15 mV

.013%+.2 mV

.042%+.26 mV .042%+.52mV

30V

.01%+1.5 mV

.02%+3 mV

.013%+1.7 mV

.026%+3.5 mV

.042%+3.9 mV .084%+7.8mV

150/300V

.01%+15 mV

.04%+30 mV

.013%+17 mV

.052%+35 mV

.042%+39 mV

.168%+78 mV

Note 300V range applies to channels 1 and 11 only.

2640A AC Voltage Measurement Specifications

A-14.

Tables A-12 to A-14 provide 2640A specifications for the ac voltage measurement
function.

Table A-12. 2640A AC Voltage General Specifications

Specification

Characteristic

Input Impedance

1 M

in parallel with 100 pF

Maximum Crest Factor

3.0 Maximum
2.0 for rated accuracy

Crest Factor Error

For non-sinusoidal input signals with crest factors
between 2 and 3 and pulse widths

100 µs, add

0.2% to the accuracy specifications.

Common Mode Rejection

80 dB minimum at dc, 50 Hz/60 Hz

±

0.1%, 1 k

imbalance, Slow Rate