2640a 4-wire resistance measurement specifications, A-12 – Fluke NetDAQ 2645A User Manual

Page 222

Advertising
background image

2640A/2645A NetDAQ

Users Manual

A-12

2640A 4-Wire Resistance Measurement Specifications

A-15.

Tables A-15 to A-17 provide 2640A specifications for the 4-wire resistance
measurement function. The 4-wire measurements use 2 input channels a decade
apart, e.g., channels 4 and 14.

Table A-15. 2640A 4-Wire Resistance Temperature Coefficient

Specification

Characteristic

Temperature Coefficient

Add 1/10th the 90 day specification per

º

C

above 28

º

C or below 18

º

C.

Accuracy at -20

°

C

Multiply the -10

°

C to + 60

°

C accuracy

specification by 2. After 1 hour warmup. For
accuracy between -10

°

C and -20

°

C, interpolate

linearly

.

Table A-16. 2640A 4-Wire Resistance Range and Resolution Specifications

Resolution

Current

Full Scale

Maximum voltage

Range

Slow

Fast

Applied

Voltage

applied by instrument

300

1 m

3 m

1 mA

300 mV

3.5V

3 k

10 m

30 m

100 µA

300 mV

3.5V

30 k

100 m

300 m

10 µA

300 mV

3.5V

300 k

1

3

10 µA

3.0V

3.5V

3 M

10

30

1 µA

3.0V

3.5V

Table A-17. 2640A 4-Wire Resistance Accuracy Specifications

Accuracy, 3

σ

+ (% input + V)

18

º

C to 28

º

C

-10

º

C to 60

º

C

Range

90 Day

1 Year

1 Year

Slow

Fast

Slow

Fast

Slow

Fast

300

.015%+20 m

.02%+80 m

.02%+50 m

.02%+120 m

.084%+126 m

.084%+336 m

3k

.02%+.3

.02%+.8

.02%+.5

.02%+1.2

.084%+1.26

.084%+3.36

30 k

.03%+3

.04%+10

.03%+5

.04%+15

.126%+12.6

.168%+42

300 k

.1%+40

. 2%+100

.1%+60

.2%+150

.42%+168

.84%+420

3 M

*

.25%+800

*

.5%+10k

*

.25%+1 k

*

.5%+1.5k

*

1.05%+3.36

2.1%+4.2 k

*

* The 3 M

range is susceptible to the absorption of humidity under extreme conditions. If the instrument is

operated normally within its specified temperature-humidity range, the 3 M

range meets its accuracy

specifications. However, if the instrument is “soaked” at 50

°

C, 90% RH the 3M

range may require 1 hour

of “dry-out” time at 25

°

C, <40% RH for each hour of “soak” time in order to achieve its specified accuracy.

Advertising
This manual is related to the following products: