2640a frequency measurement specifications, A-16 – Fluke NetDAQ 2645A User Manual

Page 226

Advertising
background image

2640A/2645A NetDAQ

Users Manual

A-16

2640A Frequency Measurement Specifications

A-20.

Tables A-22 to A-23 provide 2640A specifications for the frequency measurement
function.

Table A-22. 2640A Frequency Accuracy Specifications

Frequency Measurement Accuracy, 1 Year, -10

º

C to 60

º

C

Range

Resolution

Accuracy

±

(% input + Hz)

Slow

Fast

Slow

Fast

15 Hz to 900 Hz

0.01 Hz

0.1 Hz

0.05%+0.02 Hz

0.05%+0.2Hz

900 Hz to 9 kHz

0.1 Hz

1 Hz

0.05%+0.1 Hz

0.05%+1 Hz

9 kHz to 90 kHz

1 Hz

10 Hz

0.05%+1 Hz

0.05%+10 Hz

90 kHz to 900 kHz

10 Hz

100 Hz

0.05%+10 Hz

0.05%+100 Hz

1 MHz

100 Hz

1 kHz

0.05%+100 Hz

0.05%+1 kHz

Table A-23. 2640A Frequency Sensitivity Specifications

Frequency Measurement Sensitivity (Sine Wave)

Frequency Range

Minimum Signal

Maximum Signal

15 Hz to 70 Hz

100 mV rms

V<150/300V rms* (depends on
channel used) and V*Hz<2*10

6

70 Hz to 100 kHz

100 mV rms

20V rms

100 kHz to 200 kHz

150 mV rms

10V rms

200 kHz to 300 kHz

150 mV rms

7V rms

300 kHz to 1 MHz

Linearly increasing from 150V
rms at 300 kHz to 2V rms at 1
MHz

Linearly decreasing from 7V rms
at 300 kHz to 2V rms at 1 MHz

* 300V range applies to channels 1 and 11 only.