2645a specifications, 2645a dc voltage measurement specifications, A-17 – Fluke NetDAQ 2645A User Manual

Page 227

Advertising
background image

Specifications

2645A Specifications

A

A-17

2645A Specifications

A-21.

This section includes specifications specific to the 2644A instrument by
measurement function.

2645A DC Voltage Measurement Specifications

A-22.

Tables A-24 to A-26 provide 2645A specifications for the dc voltage measurement
function.

Table A-24. 2645A DC Voltage Measurement General Specifications

Specification

Characteristic

Input Impedance

100M

in parallel with 300 pF maximum for ranges

3V

10M

in parallel with 100 pF maximum for ranges

>3V

Normal Mode Rejection

50dB minimum at 50 Hz/60 Hz +0.1%, Slow Rate

Common Mode Rejection

120dB minimum at dc, 50 Hz/60 Hz +0.1%, 1k

imbalance, Slow Rate
80dB minimum at dc, 60 dB at 50 Hz/60 Hz +0.1%,
1k

imbalance, Medium and Fast Rates

Channel-to-Channel Crosstalk

120dB minimum Slow Rate (e.g., 30V dc on
channel 1 may cause a 30 µV error on channel 2)
80dB minimum Medium and Fast Rates (e.g., 1V
dc on channel 1 may cause a 10 µV error on
channel 2)

Temperature Coefficient

For % input: Add 1/10th the 90-day specification
per

º

C above 28

º

C or below 18

º

C.

For floor error (V): add 1/20th the 90-day
specification per

º

C above 28

º

C or below 18

º

C

Accuracy at -20

°

C

Multiply the -10

°

C to + 60

°

C accuracy specification

by 2. After 1 hour warm-up. For accuracy between
-10

°

C and -20

°

C, interpolate linearly

.

Maximum Input Voltage

The lesser voltage of:
50V dc or 30V ac rms from any input terminal to
earth
-or-
50V dc or 30V ac rms from any input terminal to
any other input terminal