A-23 – Fluke NetDAQ 2645A User Manual

Page 233

Advertising
background image

Specifications

2645A Specifications

A

A-23

2645A Thermocouple per ITS-1990 Measurement Specifications A-27.

Tables A-35 to A-36 provide 2645A specifications for the thermocouple
measurement function per ITS-1990.

Table A-35. 2645A Thermocouple General Specifications

Specification

Characteristic

Input Impedance

100 M

minimum in parallel with 150 pF

Open Thermocouple Detect

Operates by injecting a small ac signal into the
input after each measurement. A thermocouple
resistance greater than 1 k

to 10 k

is

detected as an open input.

Temperature Coefficient

To calculate Thermocouple accuracy for
temperatures between 28

º

C and 60

º

C, or -10

º

C

and 18

º

C, use a linear interpolation between

the two applicable points. e.g., if the applicable
spec at 28

º

C is .6 and the spec at 60

º

C is 1.1,

then the spec at 40

º

C is (1.1-.6)*(40-28)/(60-

28)+.6 = .5*(12.32) + .6 = .7875.

Accuracy at -20

°

C

Multiply the -10

°

C to + 60

°

C accuracy

specification by 2. After 1 hour warm-up. For
accuracy between -10

°

C and -20

°

C, interpolate

linearly.

Table A-36. 2645A Thermocouple Specifications

Accuracy

±

º

C

Thermocouple

Resolution

18

º

C to 28

º

C

-10

º

C to 60

º

C

90

Day

1 Year

1 Year

Type

Temperature

º

C

Slow

Slow

Fast

Slow

Fast

J

-100 to 80

.3

0.8

0.9

1.6

0.9

1.7

80 to 230

.2

0.7

0.8

1.4

0.9

1.5

230 to 760

.2

0.7

0.8

1.3

1.0

1.5

K

-100 to -25

.4

1.0

1.1

2.0

1.2

2.1

-25 to 120

.3

0.8

0.9

1.7

1.0

1.8

120 to 1000

.3

0.9

1.1

1.8

1.5

2.2

1000 to 1372

.3

1.2

1.5

2.3

2.0

2.9

N

-100 to -25

.5

1.4

1.5

2.8

1.5

2.9

-25 to 120

.5

1.1

1.3

2.3

1.3

2.4

120 to 1000

.4

1.0

1.1

2.0

1.2

2.1

1000 to 1300

.3

1.0

1.2

1.9

1.6

2.4