2645a frequency measurement specifications, A-25 – Fluke NetDAQ 2645A User Manual

Page 235

Advertising
background image

Specifications

2645A Specifications

A

A-25

2645A Frequency Measurement Specifications

A-28.

Tables A-37 to A-38 provide 2645A specifications for the frequency measurement
function.

Table A-37. 2645A Frequency Accuracy Specifications

Frequency Measurement Accuracy, 1 Year, -10

º

C to 60

º

C

Range

Resolution

Accuracy + (% input + Hz)

Slow

Fast

Slow

Fast

15 Hz to 900 Hz

0.01 Hz

0.1 Hz

0.05%+0.02 Hz

0.05%+0.2 Hz

900 Hz to 9 kHz

0.1 Hz

1 Hz

0.05%+0.1 Hz

0.05%+1 Hz

9 kHz to 90 kHz

1 Hz

10 Hz

0.05%+1 Hz

0.05%+10 Hz

90 kHz to 900 kHz

10 Hz

100 Hz

0.05%+10 Hz

0.05%+100 Hz

1 MHz

100 Hz

1 kHz

0.05%+100 Hz

0.05%+1 kHz

Table A-38. 2645A Frequency Sensitivity Specifications

Frequency Measurement Sensitivity (Sine Wave)

Frequency Range

Minimum Signal

Maximum Signal

15 Hz to 70 kHz

100 mV rms

30V rms

70 kHz to 100 kHz

100 mV rms

20V rms

100 kHz to 200 kHz

150 mV rms

10V rms

200 kHz to 300 kHz

150 mV rms

7V rms

300 kHz to 1 MHz

Linearly increasing from 150V
ms at 300 kHz to 2V rms at 1
MHz

Linearly decreasing from 7V
rms at 300 kHz to 2V rms at 1
MHz

Advertising