Appendix D

RTD Linearization

Introduction

D-1.

This appendix discusses Fixed-385 and Custom-385 RTD measurements with
NetDAQ.

Fixed-385 RTD

D-2.

The Fixed-385 RTD linearization uses segmented polynomials to approximate the
Callendar Van-Dusen equation for a 385 RTD. R

0

may be specified via NetDAQ

Logger but

α

,

δ

, and

β

are fixed. These fixed coefficients are the modified DIN

43760 coefficients that conform to ITS90:

α

= 0.00385055

δ

= 1.49978574

β

= 0.10863383

The accuracy specifications of Fixed-385 measurements with a 385 RTD that
matches the above coefficients are shown in Table A-34 in Appendix A.