RTD Linearization

Custom-385 RTD

D

D-3

Using the Custom-385 RTD with Other Platinum RTDs

D-4.

Although the Custom-385 RTD linearization uses a

β

and

δ

from a 385 RTD,

other RTD types can be used. The Custom-385 linearization will compensate for
the change in

α

, but errors will be introduced due to the difference in the

δ

and

β

coefficients of the RTD and the fixed coefficients. Figure D-2 shows the accuracy
envelope for a 375 RTD. Because of the error introduced by the approximation
and the difference in

δ

it is recommended that the use of the 375 RTD be limited

to temperatures between -100

°

C and 300

°

C. Figure D-3 shows the accuracy

envelope for a 391 RTD and Figure D-4 shows the accuracy envelope for a 392
RTD. Due to errors introduced by the approximation and the difference in

δ

it is

recommended that the use of the 391 RTD and 392 RTD be limited to
temperatures between -100

°

C and 600

°

C.

2645A
2640A

-100

-50

0

50

100

150

200

250

300

Temperature - Degrees C

Accuracy Envelope: Custom RTD, 375 Probe

1 Year Slow: 18-28 Degrees C

Alpha: 0.00375 Delta: 1.605 Beta: 0.16

-2.0

-1.5

-1.0

-0.5

0.0

0.5

Error - Degrees C

1.0

Figure D-2. 375 RTD

Advertising