WARNING:

IF VOLTAGES GREATER THAN 50V (2640A ONLY) ARE TO BE MEASURED, A
SEPARATE EARTH GROUND MUST BE ATTACHED TO THIS REAR PANEL
GROUND CONNECTOR WHEN THE INSTRUMENT IS OPERATED FROM BATTERY
POWER.

Line Cord (AC Operation)

External Battery (DC Operation)

+ –

Ground for 50-ohm
Termination Ground Lug.

+

WARNING

: TO AVOID ELECTRICAL SHOCK, DISCONNECT LINE CORD BEFORE REMOVING COVER

9-16V

DC PWR

DIGITAL I/O

MA TO TI

0

1

2

3

4

5

6

7

+30V

ALARM/TRIGGER I/O

ETHERNET

NOT FOR CONNECTION TO

PUBLIC TELEPHONE SYSTEMS

MEETS VFG 243 / 1991

OVERVOLTAGE CATEGORY II PER IEC 1010-1

SERIAL PORT

MODEL

: 2640A / 41A

2645A / 46A

ON / OFF

XMT RCV

LK

107-264V

50/60 Hz
15VA

Figure 2-2. Connecting the Instrument to a Power Source

P

Warning

To avoid shock hazard when powering the instrument
with ac power, connect the power cord to a receptacle
with an earth ground.

AC Power

2-6.

Plug the line cord into the connector on the rear of the instrument as shown in
Figure 2-2. The instrument operates on any line voltage between 107 and 264V ac
(45 to 65 Hz) without adjustment. Fluke warrants the instrument to meet
specifications only at 50 Hz and 60 Hz operation. Power consumption is a
nominal 15 watts. Be sure the line cord ground terminal connects to an earth
ground.

P

Warning

To avoid shock hazard when powering the instrument
with dc power, connect the instrument ground terminal to
an earth ground.