Reviewing and Setting the General Network Socket Port

2-33.

Perform the procedure in Figure 2-25 to review or set the general network Socket
Port (1024 to 65535). The default is 04369. In order to communicate with each
other, a host computer and an instrument must use the same socket port number.
(See Figure 2-25 for examples.)

General Network Socket Port Enter the Socket Port supplied by your
network administrator.

COMM

Press the COMM key to review the network settings, or press and hold the

COMM key for 3 seconds to set the network settings.

Press the up/down arrow keys until nEt (Network) appears in the primary

display (comm appears in the secondary display).

ENTER

Press the ENTER key. nEt appears in the secondary display and ISo (isolated
network) or gEn (general network) is in the primary display.

To set the network type, press the up/down arrow keys to

select gEn.

ENTER

Press the ENTER key. (Pressing any other function key will cancel set

operations.) This displays the current Socket Port.

To set the socket port, press the left/right arrow keys to select the desired

digit position (highlighted).

Press the up/down arrow keys to select the desired number, 0 to 9, for the

positioned Port digit. In this manner, select all Port digits.

ENTER

Press the ENTER key. This displays the first digit of the Internet Protocol
address segment IP:0.

ENTER

Press the ENTER key to enter the settings and exit the procedure. (Pressing any

other function key will cancel set operations.).

Figure 2-25. Reviewing and Setting the General Network Socket Port