28

SYMPTOM

High operating pressure

Excessive pressure drop

[over 50 psig (3.4 barg)]

(high primary pressure - low

final pressure)

Water flowing when machine is

turned OFF

Declining rejection

(high permeate conductivity)

POSSIBLE CAUSES

Recycle or concentrate valve

plugged

Inaccurate pressure gauge

Restricted or reduced permeate

flow rate

Restricted flow after pump

outlet

Telescoped membrane element

covering membrane element

housing outlet port

Severely fouled or dirty

membrane elements

Inlet solenoid valve not closing

or seating properly

Dirty or fouled membrane ele-

ments

REMEDIES

Disassemble the plumbing to the

recycle valve and remove for-

eign particles. Check the con-

centrate valve stem.

Replace or calibrate the gauge as

required.

See possible causes for low per-

meate rate.

Check for blockage of the con-

centrate flow at the inlets and

outlets of membrane element

housings.

Ensure that the anti-telescoping

(ATD) is located properly on the

membrane element.

Flush the machine, then clean it

with detergent.

Clean or replace the valve.

Clean the membrane elements

with detergent immediately.

Water must not pass through the

inlet when the machine is OFF.

Flush and clean the membrane

elements.

TROUBLESHOOTING GUIDE