3. Installation

3.5 Installation of
the flow cell

Monitor UV-1 accepts flow cells for Standard Chromatography, FPLC

and industrial applications.

Product Material of Path Total dead Illuminated Pressure Application
Code No. wetted parts length volume volume limit area

S-2 Fluoro-plastic, 2 mm 80 µl 2 µl 0.3 MPa Standard
19-4840-02 optical quartz (3 bar) Chromatography

HR-10 Fluoro-plastic, 10 mm 24 µl 8.7 µl 1.0 Mpa FPLC, Standard
19-6254-02 optical quartz,titanium (10 bar) Chromatography

3 mm Fluoro-plastic, 3 mm 50 µl 3 µl 1.0 MPa Preparative FPLC,
19-2503-02 optical quartz (10 bar) Standard Chromatography

10 mm Fluoro-plastic, 10 mm 250 µl 8.7 µl 1.0 MPa Standard
19-2504-02 optical quartz (10 bar) Chromatography

5 mm Silicon rubber, 5 mm 0.2 MPa* Industrial scale
Industrial polypropylene, (2 bar)

19-4510-02 optical quartz

* Flow rate 300 I/h at 0.1 MPa

All flow cell are mounted in the holders, ready to be installed directly

into the cell housing. Release the locking knob (Fig. 4:6) by turning it

against the direction of the arrow and insert the flow cell in its holder

into the optical unit. Lock it in position by turning the locking knob

fully in the direction of the arrow.
For more information please refer to the instruction sheet supplied

with the respectively flow cell.

Fig. 7.

S-2 flow cell

Fig. 8.

HR-10 flow cell

Fig. 9.

3 mm flow cell

Fig. 10.

10 mm flow cell

Fig. 11.

Industrial flow cell