Trouble-shooting – GE UV-1 User Manual

Page 23

Advertising
background image

6. Trouble-shooting

6. Trouble-shooting

The UV-1 Monitor has been designed for trouble-free use. If good chromatographic practice is

followed, very little difficulty should be experienced. Clean optical surfaces are essential if low noise

levels are to be maintained. The following check list of the most frequent problems is meant to be a

guide in trouble-shooting. If the checks in this section are executed and the UV-1 still does not work

properly, consult your local GE Healthcare representative.

Symptom Cause Remedy

Pilot light does not light 1. Mains cord not plugged in Check that mains cord is plugged in

2. Fuse blown Replace fuse. If fuse blows again

immediately consult

GE Healthcare

representative

3. No voltage at mains socket Check by plugging table lamp in

Pilot light on, no recorder 1.

response

2.

3.
4.

5.

6.

7.

8.

Optical unit not connected

to control unit

Control unit not connected

to recorder

Recorder not operating

Recorder zero not set correctly

Recorder range incorrect

Wrong filter

Filter not pushed in fully

Shutter in place

Check that connecting cable is
plugged in

Check connection between output

terminals and recorder

Check recorder function
Zero recorder

Set recorder range to 10 mV

Check that the filter corresponds to

converter or aperture

Push in filter

Remove shutter and insert converter

or aperture as appropriate

Excessive noise 1.

2.

3.

4.

5.

6.

7.

8.

8.

9.

Poor ground contact

Excess noise on mains supply

Recorder connections

incorrect

Recorder range incorrect
Dirty cell

Deposits on optical surfaces

Solvent with high UV absorption

Lamp not warmed up

Bubbles passing through the cell

Aging lamp

Check contact to ground

Use alternative power source or
remove source of disturbance

Check connection between

output terminals and recorder

Set recorder on 10 mV range

Clean cell

Clean filter and converter

Relocate optical unit in a clean environment

Change to a more suitable solvent
2 h warm-up, in cold room 12 h

De-gas solvent. Check for leaks

Check lamp and replace if necessary

Excessive baseline drift 1.

2.

3.

4.

5.

Variable absorbance gradient Compensate by use of reference cell
Contaminated solvent Use fresh solvent. Check that plastic tubing

does not leak UV absorbing substances

Large variation in ambient Relocate optical unit or remove

temperature source of temperature change

Instrument warm-up Allow 2 hours warm-up

Condensation forming in Flush reference cell with dry gas or

empty reference fill it with the appropriate solvent

Long term noise, 1. Variations in ambient temperature Relocate optical unit or protect

often regular waves in especially in cold room from draught
recorder reponse 2. Flow rate variations Check pump system and column packing

3. Poor ground contact Check contact to ground
4. Bubbles passing cell De-gas solvent. Check for leaks

5. Dirty cell Clean cell
6. Deposits on optical surfaces Clean filter and converter

Relocate optical unit in a clean environment

23

Advertising