7. Technical Specifications

7.

Technical Specifications

Wavelength range

Lamp

Filters

Operating modes
Full scale ranges

Noise

Linearity

Temperature drift

Long term drift

Time constant
Recorder output
Environment

Power consumption
Power supply, voltage

frequency

Dimensions (LxWxD)

Weight

254 nm, 280 nm and 405 nm
Hg lamp: for 254, 280, and 405 nm

Life time lamp: 8000 hours

converter: 2000 hours

Interference. Stray light maximum 0,1 %

at 254 nm, maximum 0.8% at 280 nm
AU or Transmission
0.01, 0.02, 0.05, 0.1, 0.2, 0.5, 1 or 2 AUFS
4 x 10-5 AU peak to peak maximum at
254 nm (dry cell)

2 x 10-4 AU peak to peak (typical at
254 nm in flowing liquid)

5 x 10-4 AU peak to peak (typical at
280 nm flowing liquid)
At 254 nm, better than ± 3% to 2 AU

At 280 nm, better than ± 5% to 1 AU
2 x 10-4 AU/°C typical with dry cell at
254 nm

x 10-3 AU/°C typical with dry cell

at 280nm
1 x 10-4 AU/h at 254 nm, constant
temperature after 2 hours warm-up

4 x 10-4 AU/h at 280 nm, constant
temperature after 2 hours warm-up

1.5 s to 90% of FSD at all ranges

0-10 mV
0 to +40 °C, 20-95 % relative humidity,

84-106 kPa (840-1060 mbar) atmospheric

pressure
20 VA

100/120/220-230/240

5

0-60 Hz

Control unit: 180x145x75 mm

Optical unit: 180x145x75 mm
Control unit: 1.6 kg Optical unit: 1.7 kg

EMC standards This product meets the requirement of the EMC Directive 89/336/EEC through the harmonized standards

EN 50081-2 (emission) and EN 50082-1 (immunity)

Note: This is a class A product. In a domestic environment this product may cause radio interference in which

case the user may be required to take adequate actions.

Note: The declaration of conformity is valid for the instrument when it is:

used in laboratory locations

used in the same state as it was delivered from

GE Healthcare Bio-Sciences AB

except for alteration described in the

User Manual

used as “stand alone” unit or connected to other CE labelled

GE Healthcare

products or other products as

recommended.

Safety standards This product meets the requirement of the Low Voltage Directive (LVD) 73/23/EEC through the harmonized

standard EN 61010-1.