4 interference filters, 5 other optical surfaces, 6 instrument housing – GE UV-1 User Manual

Page 20: 7 lamp and optical system test

Advertising
background image

8. Insert the new cell, being careful not to touch either of the optical

surfaces. See that the inlet and outlet ports are correctly aligned.

9. Retighten the screw washers with the tool provided. Do not use

excessive force.
10. Check for leakage.
11. Replace the black cover.

For optimum performance, it is essential that the interference filters

are clean and free of any particulate material. Do not touch the inter-

ference filters. Should the filter become contaminated with dust,finger

prints or oil, proceed as follows:
- Carefully take out the filter without touching or scratching the

surface.
- Use lens cleaning tissue dipped in ethanol to gently clean both

sides of the filter surface.
- Place the clean filter back to the UV-1 or its box.
- Interference filters should never be exposed to temperatures

above 60

°

C

Clean all other optical surfaces by wiping the surface with clean, lint-

free cloth, moistened with carbon tetrachloride, ethanol, or another

suitable pure solvent.

Wipe the instrument regularly with a damp cloth. Let the instrument

dry completely before use.

Before performing this test see that the filter is clean, that the cell is

clean and free of bubbles or particles and that the cell holder is

correctly inserted and locked in position.
1.

Insert the aperture and filter for 254 nm operation or the

converter and filter for 280 nm operation.
2.

Set AU / %T to % T.

3.

Set the recorder to the 100 mV range.

20

5. Maintenance

Reference out

Sample out

Reference in

Sample in

5.4 Interference
filters

5.5 Other optical
surfaces

5.6 Instrument
housing

5.7 Lamp and
optical system
test

Fig. 13.

Flow cell interior showing tubing connections.