Metrologic Instruments StratosS MS2220 User Manual

Page 54

Advertising
background image

48

S

CALE

O

PERATION

C

ALIBRATION


Bar Code Calibration Procedure without Remote Display*

6. Make sure there is no load on the scale platter. The white LEDs will be used to indicate the current step in the calibration

process.

7. Calibrate a zero load. The white LEDs will blink once periodically. Wait 8 to 10 seconds for scale stability, and then

press the Multi-Function Button once. After the Multi-Function Button has been pressed, there will be a short delay
followed by 1 short beep indicating that the zero load value has been stored.

Figure 51.

8. Place

a

FULL Load on the center of the scale platter. The white LEDs will blink 3 times periodically. Wait at least 3

seconds for scale stability, and then press the Multi-Function Button once. After the Multi-Function Button has been
pressed, there will be a short delay followed by 3 short beeps indicating that the FULL Load value has been stored.

Figure 52.

9. Remove the entire load from the scale platter. The white LEDs will blink 5 times periodically. Wait at least 3 seconds for

scale stability, and then press the Multi-Function Button once. After the Multi-Function Button has been pressed, there
will be a short delay followed by 5 short beeps indicating that the zero load value has been stored.

Figure 53.

10. Calibration process is now complete. The scanner will automatically restart, and beep one time, in approximately

12 seconds.

11. Verify the calibration was successful with the calibrated weight. If the scale display does not accurately show the

calibrated load weight correctly, restart this procedure from Step 1 on page 45.


12. The unit’s calibration must now be verified as required by state and/or local Weight and Measures regulations

(starting on Page 49).

* These procedures are for MS2220 StratosS Scanner/Diva Scale models only

.

Advertising