A r m s – Invacare Allegro 1061057 User Manual

Page 25

Advertising
background image

25

A
R

M

S

FIGURE 12 - REPLACING/REPOSITIONING THE

LOCKING MECHANISM IN THE FLIP-UP HALF ARM

ARM PAD DEPTH ADJUSTMENT/
REPLACEMENT - CANTILEVER ARMS
(FIGURE 13)

Adjustment

1. Remove the phillips screw from the rear of the arm-

rest pad and self-taping screw if FULL LENGTH ARM
PADS
.

2. Depending on the desired arm pad depth, reposi-

tion the cantilever slide tube to one (1) of five (5)
positions for DESK LENGTH ARM PADS and into
the FIRST adjustment hole for the FULL LENGTH
ARM PADS
.

3. Reattach the arm pad/cantilever slide tube to the

arm tube with existing hardware.

4. Repeat for the opposite side, if necessary.

Replacement

1. Remove the phillips screws from the armrest pad.

2. Replace with NEW armrest pad.

3. Secure with existing hardware.

FIGURE 13 - ARM PAD DEPTH ADJUSTMENT/

REPLACEMENT - CANTILEVER ARMS

Phillips Screw
(Rear of Armrest
Pad)

Cantilever Arm
Self-Tapping Screw

Cantilever Slide Tube

NOTE: Only these Adjust-
ment Holes can be used.

NOTE: This Phillips Screw only needs to be re-
moved when Armrest Pad is replaced.

Full Length

Arm Pads

NOTE: This Phillips Screw only needs to be re-
moved when Armrest Pad is replaced.

ARMS

PROCEDURE 2

3. Refer to FIGURE 6 to determine the mounting hole

in the adjustment plate that will be used to corre-
spond to the back angle.

NOTE: Back angles of 105

o

and 110

o

will use the same

arm adjustment plate mounting holes.

4. Securely tighten the locking pin and washer to the

adjustment plate with a locknut.

5. Repeat STEPS 1-4 for the opposite side, if neces-

sary.

REPLACING/REPOSITIONING THE LOCKING
MECHANISM IN THE CANTILEVER ARM
(FIGURE 12)

1. Move the cantilever arm up and out of the way.

2. Remove the actuator from the locking mechanism.

CAUTION

The locking mechanism is spring loaded. Place your
free hand over the locking mechanism to prevent
the parts from springing out of the cantilever arm.

3. Slowly let the locking mechanism and spring slide

out of the cantilever arm.

NOTE: Inspect the spring and replace if necessary.

4. Slide the new locking mechanism and spring into

the cantilever arm.

5. Position the angled portion of the locking mecha-

nism in one (1) of two (2) ways:

Angled Portion Facing Up - Arm will flip UP
Angled Portion Facing Down - Arm will flip DOWN

6. Use Loctite 242 and securely tighten the actuator

into the locking mechanism.

7. To lock the cantilever arm, push down until there is

an audible click.

8. Pull up on the cantilever arm to make sure it is locked

in place.

Angled Portion of

Locking Mechanism

Actuator

(Apply Loctite 242)

Locking

Mechanism

Cantilever Arm

Spring

Desk Length

Arm Pads

Phillips Screw
(Rear of Armrest
Pad) Cantilever Slide Tube

Advertising
This manual is related to the following products: