B a c k – Invacare Allegro 1061057 User Manual

Page 34

Advertising
background image

34

FIGURE 16 - BACK ANGLE PLATE MOUNTING

HOLES FOR CORRESPONDING BACK

ANGLES

2. Remove the rear hex bolt, washer and locknut that

secure the back angle plates to the wheelchair frame
(FIGURE 15).

3. Refer to FIGURE 16 to determine the mounting hole

in the back angle plates for the necessary back angle.

4. Reinstall the rear hex bolt, washer and securely tighten

with locknut. Refer to FIGURE 15 for correct hard-
ware orientation.

5. If the wheelchair is equipped with cantilever arms,

adjust the arms to keep them parallel to the ground/
floor. Refer to USING/INSTALLING/HEIGHT AD-
JUSTMENT/CORRESPONDING ARM ADJUST-
MENT TO BACK ANGLE - CANTILEVER ARMS
in
PROCEDURE 2 of this manual.

NOTE: If the wheelchair has become less stable af-
ter changing the back angle, reposition the rear
wheels. Refer to ADJUSTING THE WHEELBASE
LENGTH in PROCEDURE 6 of this manual.

REPOSITIONING THE BACK (CHANGING
SEAT DEPTH)

WARNING

The seat height, seat depth, back angle, seating sys-
tem, size and position of the rear wheels, as well as
the user condition directly relate to the stability of
the wheelchair. Any change to one (1) or any combi-
nation of the seven (7) may cause the wheelchair to
decrease in stability. These adjustments MUST be
performed by an authorized dealer or qualified tech-
nician.

Non-Recliners (FIGURE 17)

1. Determine the necessary back position for the de-

sired seat depth of the wheelchair. Refer to CHANG-
ING SEAT DEPTHS
in PROCEDURE 10 of this
manual.

2. Remove the rear wheels from the wheelchair. Refer

to REMOVING/INSTALLING REAR WHEELS in
PROCEDURE 6 of this manual.

3. Remove the four (4) hex bolts, washers and locknuts

that secure the four (4) back angle plates to the wheel-
chair frame.

NOTE: Note the back angle before removing the hex
screws for reinstallation.

4. Line up the mounting holes in the back angle plates

with the mounting holes determined in STEP 1.

80

O

Back Plate

95

O

Back Plate

85

O

Back Plate

100

O

Back Plate

90

O

Back Plate

105

O

Back Plate

Back Plate

110

O

CHANGING THE BACK ANGLE

WARNING

The seat height, seat depth, back angle, seating sys-
tem, size and position of the rear wheels, as well as
the user condition directly relate to the stability of
the wheelchair. Any change to one (1) or any combi-
nation of the seven (7) may cause the wheelchair to
decrease in stability. These adjustments MUST be
performed by an authorized dealer or qualified tech-
nician.

1. Loosen, but do not remove the front hex bolt

that secures the back angle plate to the wheel-
chair frame (FIGURE 15).

PROCEDURE 3

BACK

B
A
C
K

Wheelchair

Frame

Locknut

Washer

Rear Hex

Bolt

Front Hex Bolt (Loosen, but

do not Remove)

Back Angle

Plate

FIGURE 15 - CHANGING THE BACK ANGLE