Foreword

Notes on Use

• This instrument should be used under the following operat-

ing conditions.

Temperature of 0 to 40°C (32 to 104°F), relative humidity

of 30 to 85%, with no condensation.

Atmospheric pressure of 700 to 1060 hPa (altitude of

400 to 3000m).

• This is a precision instrument. To avoid the possibility of it

being damaged, the instrument should not be dropped nor

should heavy objects be placed on top of it.

• Do not expose this instrument to rain and water.

• This instrument has no alarm function. Do not use this in-

strument if an alarm function is necessary.

• This instrument is designed for use on adults. When using

the instrument on infants or babies, please note that the de-

signed measurement accuracy cannot be guaranteed.

• Do not use the instrument on a point where there is exces-

sive vibration. In addition, to avoid excessive impact on the

instrument, it should be handled gently. Failure to observe

this may cause breakdown.

• A maximum of 300 hours of measurement data can be stored,

or the maximum storable number of data files is 399 files. If

measurements are continued after the maximum number of

storable data has been reached, the new data will overwrite

the oldest data. To keep data, be sure to download it before

the memory becomes full.

• Accurate measurement data may not be obtained in the fol-

lowing cases.

When body movement is excessive

When the probe is not attached properly

When blood circulation at the measuring point is poor (due to

pressure on arm/finger, peripheral circulatory insufficiency)

When ambient light (e.g. panel light, fluorescent light, in-

frared heat lamp, direct sunlight) is too strong