Foreword, When a mobile telephone is used during measurement, When finger nails are polished – Konica Minolta Oxygen Saturation Monitor PULSOX-300i User Manual

Page 11

Advertising
background image

9

Foreword

Difference from the SaO

2

value to displayed value

<Table 1>

<Table 2>

SaO

2

HbCO

SaO

2

Methemoglobin

1%

5%

10%

1%

5%

10%

50

50

70

70

90

90

100

100

Measured value is always
lower than the actual value.

Displayed value is sometimes
higher than the actual value if

SaO

2

is around 50%.

-0.1

-0.7

-1.5

+0.2

+1.3

+3.2

-0.2

-0.8

-1.6

-1.5

-6.0

-9.6

-0.1

-0.7

-1.5

-0.6

-2.3

-3.2

-0.2

-0.8

-1.7

-1.8

-7.5

-12.2

When influenced electromagnetically by other electronic de-

vices (e.g. near electrical appliances like TV, medial device)

When a mobile telephone is used during measurement

When influenced by abnormal hemoglobin like carbon

monoxide hemoglobin (HbCO) and methemoglobin

When pigments such as cardiogreen, intravascular dyes

and indocyanine greenare present in the blood

When finger nails are polished

Tables 1 and 2 below show the errors which may occur due to

these hemoglobins.

• Do not immerse the probe, the extension cable or the con-

necter of PULSOX-300i in disinfectant. Doing so may cause

a malfunction.

• When using the Extension Cable EC-300, fit and secure ca-

bles so they are not too loose or too tight.

Before connecting to a personal computer, remove the Pulsox

from the patient.