Measurement principle – Konica Minolta Oxygen Saturation Monitor PULSOX-300i User Manual

Page 36

Advertising
background image

34

Appendix

¢

Spectral Absorption of Hb and HbO

2

Absorption coef

ficient

Wavelength

Red light

Infrared light

Measurement Principle

This instrument measures the oxygen saturation (SpO

2

) in arte-

rial blood and pulse rate continuously in a non-invasive method

by applying optical principles.

With this instrument, SpO

2

is defined by the following equation.

where

C (Hb):

Concentration of reduced hemoglobin

C (HbO

2

): Concentration of oxyhemoglobin

The light-absorption characteristics of reduced hemoglobin (Hb)

are very different from those of oxyhemoglobin. This instrument

measures the changes in the absorption of red and infrared

lights passing through the tissue to determine the SpO

2

of the

blood. Thus, this method is free from the effects of skin color,

muscles, bones, and veins.