TPS1200

Check & Adjust

3.

TOOLS Tilting-Axis Adjustment I

Aim the telescope accurately at a target
at about 100 m distance or less if not
possible. The target must be positioned
at least 27°/30 gon above or beneath the
horizontal plane.
The procedure can be started in any
telescope face.

Step

Description

TPS12_024a

V=90

°

+ 27°

- 27°