LEICA TPS1200 User Manual

Page 98

Advertising
background image

98

TPS1200

Check & Adjust

5.

MEAS (F1) to measure and to continue to the next screen.

Motorised instruments change auto-
matically to the other face.

Non-motorised instruments guide to
the other face.

)

The fine pointing has to be

performed manually in both faces.

6.

TOOLS Combined II

MEAS (F1) to measure the same target in the other face and to calculate
the instrument errors.

)

If one or more errors are bigger than the predefined limits, the procedure
has to be repeated. All measurements of the current run are rejected and
none of them is averaged with the results from previous runs.

7.

TOOLS Adjustment Accuracy

Step

Description

TPS12_025

180°

180°