5 adjustment of the circular level, To "5.5 adjustment of the circular level – LEICA TPS1200 User Manual

Page 106

Advertising
background image

106

TPS1200

Check & Adjust

5.5

Adjustment of the Circular Level

On the instrument
step-by-step

Step

Description

1.

Level up the instrument in advance with the electronic level, assuming that
the electronic level is correctly adjusted. SHIFT F12 to access STATUS
Level & Laser Plummet
.

2.

The bubble must be centered. If it extends beyond the circle, use the allen
keys supplied to centre it with the adjustment screws. Turn the instrument
slowly 200 gon (180°). Repeat the adjustment procedure if the bubble
does not stay centered.