Locked setting lock, Locken setting lock enabled – Laser beta lasermike LS8000-3 User Manual
Page 149

LaserSpeed 8000-3 Instruction Handbook
RS-232/RS-422 Communication
Part No. 93463 / Drawing No. 0921-01561
Page 149 of 221
Revision A (Sep 2007)
*LOCKED
Setting Lock
*LOCKED?<CR>
Read
Response
Format:
*LOCKED=<N><CR>
*LOCKED=<N><CR>
Set
<N>
is either 0 or 1. This setting controls the Settings Lock. In order to prevent accidental
changes to the LS8000-3's configuration settings, you must explicitly unlock the settings
before any changes can be made.
Unlock Command: *LOCKED=0<CR>
Relock Command: *LOCKED=1<CR>
After unlocking the settings with *LOCKED=0<CR>, you can change settings. If you try to
change setting without unlocking the LS8000-3, the changes will be ignored.
The LS8000-3 will automatically re-arm the Setting Lock after 1 minute.
*LOCKEN
Setting Lock Enabled
*LOCKEN?<CR>
Read
Response
Format:
*LOCKEN=<N><CR>
*LOCKEN=<N><CR>
Set
<N>
is either 0 or 1. If you are manually changing many settings over a period of time, the
Settings Lock can a become nuisance, so you may wish to disable it. Before the Settings
Lock can be disabled, however, the settings must be unlocked:
To disable Settings Lock:
*LOCKED=0<CR>
*LOCKEN=0<CR>
To re-enable Settings Lock:
*LOCKEN=1<CR>
The Settings Lock should be re-enabled after configuration is completed, and the LS8000-3
is to be placed into normal operation.