Autoaddr udp power-on data destination ip address, Autoport udp power-on data destination port – Laser beta lasermike LS8000-3 User Manual
Page 163

LaserSpeed 8000-3 Instruction Handbook
RS-232/RS-422 Communication
Part No. 93463 / Drawing No. 0921-01561
Page 163 of 221
Revision A (Sep 2007)
*AUTOADDR
UDP Power-On Data Destination IP Address
*AUTOADDR?<CR>
Read
Response
Format:
*AUTOADDR=<N><CR>
Set
*AUTOADDR=<XXX.XXX.XXX.XXX><CR>
Where <XXX.XXX.XXX.XXX> is an IP Address, such as "192.168.10.246". This
command configures the destination IP address for UDP output data. If the gauge is not
configured to output UDP data on start-up, then this setting is not used.
*AUTOPORT
UDP Power-On Data Destination Port
*AUTOPORT?<CR>
Read
Response
Format:
*AUTOPORT=<N><CR>
Set
*AUTOPORT=<N><CR>
Where <N> is an IP port number, such as 1234. This command configures the destination
Port for UDP output data. If the gauge is not configured to output UDP data on start-up, then
this setting is not used. This port should be set to a value higher than 1024 so that it doesn't
interfere with standard network interfaces.