Applying Hand/Foot Sensors

Neonatal hand or foot, for patients between 1kg and 4kg.

A-Series: M1193A

Special Edition (SE) A-Series: M1193AN

Applying Ear-Clip Sensors

Adult or pediatric ear, for patients more than 40kg.

A-Series: M1194A, or

Special Edition (SE) A-Series: M1194AN

Step

Applying the Hand/Foot Sensor

1

Position the sensor over the hand or foot with optical components
opposite each other.

2

Holding sensor, stretch strap slightly (not more than 2.5cm (1.0”).

3

Insert stretched strap into slot, hold it there while threading end
through latch. If strap is too long, thread it through second latch.

4

Plug sensor into instrument (or into adapter cable if needed).

5

Inspect and change the sensor application site periodically.

Step

Applying the Ear-Clip Sensor

1

To improve perfusion, massage or warm the ear lobe.

2

Clip probe onto fleshy part of the lobe. Plastic fixing mechanism
minimizes artifact generated by patient motion. Do not position
probe on cartilage or pressed against the head.

3

Plug sensor into instrument (or into adapter cable if needed).

4

Inspect and change the sensor application site periodically.