Technical information

HeartStart FRx 861304 Defibrillator specifications

The specifications provided in the following tables are nominal values. Additional
information can be found in the Technical Reference Manual for HeartStart
Defibrillators, located online at www.medical.philips.com.

Physical

Environmental

category

specifications

size

2.4” H x 7.1” D x 8.7” W
(6 cm H x 18 cm D x 22 cm W).

weight

Approximately 3.5 lbs (1.6 kg) with battery and pads installed.

pads compatibility

HeartStart SMART Pads II 989803139261
(In an emergency or during use, HeartStart DP2/DP6 pads may be used.
However, the FRx should not be stored with DP2/DP6 pads installed, as the
daily self-test will not give a “pass” result and the device will chirp.)

category

specifications

temperature and

relative humidity

Operating and standby (battery installed, pads connected):
32° to 122° F (0° to 50° C);
10% to 75% RH (non-condensing).
Storage/shipping (with battery and pads case):
-4° to 140° F (-20° to 60° C) for up to 1 week;
0% to 85% RH (non-condensing) for up to 2 days, thereafter 65% RH maximum

altitude

0 to 15,000 feet (0 to 4,572 m).

shock/drop abuse

tolerance

Withstands 1 meter drop on any edge, corner, or face of the device onto
masonry surface.

vibration

Operating: meets MILSTD 810F Fig. 514.5C-17, random.
Standby: meets MILSTD 810F Fig. 514.5C-18, swept sine (helicopter).