Patient treatment protocol options

parameter

settings

default

default description

“call EMS” voice

reminder timing

• At power on (when the

user turns on the FRx)

• At power on and at the

start of the first patient
care pause

• At the start of the first

patient care pause

• No reminder

At the start of
the first
patient care
pause.

Provides a voice reminder to
make sure emergency medical
services have been called, at
the start of the first patient
care pause.

shock series

1, 2, 3, 4

1

The automatic protocol pause
for CPR is activated each time
a shock is delivered.*

During the protocol pause,
the FRx does not perform
rhythm analysis.

The length of the protocol
pause after a shock series is
completed is determined by
the protocol pause timer
setting.

shock series interval

(minutes)

1.0, 2.0,

(infinity)

1.0

A delivered shock must occur
within 1 minute of the
previous shock to be counted
as part of the current shock
series.

NOTE: This parameter is only
applicable when the shock series
is not configured to the default 1
shock.

* A shock series begins when a shock is delivered after the FRx is turned on. A new shock series begins after a protocol

pause. If shock series is configured for 2 or more, a new shock series also begins if the time since the previous shock
exceeds the shock series interval setting.