Philips 861304 User Manual

Page 42

Advertising
background image

34

Ph
ilip

s M
e

di

ca

l Sy
st

e

m

s

HEARTSTART FRx 861304 OWNER’S MANUAL

AP

P

E

ND

IC

ES

Controls and indicators

sealing

Meets IEC 529 class IPx5 for jetting water and class IP5x for solid objects (dust
protected).

ESD/EMI (radiated and

immunity)

See Appendix F.

aircraft: method

Meets RTCA/DO-160D:1997 Section 21 (Category M - Charging).

category

specifications

category

specifications

controls

Green On/Off button

Blue i-button

Orange Shock button

Optional Infant/Child Key accessory

indicators

Ready light: green, blinks when the defibrillator is in standby mode (ready for
use); solid when the defibrillator is being used.

i-button: blue, flashes when information is available, on solid during patient care
pause.

Caution light: flashes when the defibrillator is analyzing, comes on solid when
the defibrillator is ready to deliver a shock.

Shock button: orange, flashes when the defibrillator is charged and ready to
deliver a shock.

Pads Placement LEDs: flash when FRx is turned on; off once pads are placed on
patient. Also operates with Infant/Child Key inserted to indicate pads placement
on infants and children under 55 pounds (25 kg) or 8 years old.

audio speaker

Provides voice instructions and warning tones during normal use.

beeper

Provides chirps when troubleshooting is needed.

status indicator

Status indicator LCD displays device readiness for use.

low battery detection

Automatic during daily periodic self-testing.

low battery indicator

Alarm chirps and flashing blue in-button.