Accessories specifications

HeartStart SMART Pads II 989803139261

non-shockable rhythm — all

other non-shockable

rhythms

c

450

specificity >95%

(88%)

a. From Philips Medical Systems Heartstream ECG rhythm databases.
b. American Heart Association (AHA) AED Task Force, Subcommittee on AED Safety & Efficacy. Automatic External Defibrillators for Public

Access Use: Recommendations for Specifying and Reporting Arrhythmia Analysis Algorithm Performance, Incorporation of New
Waveforms, and Enhancing Safety. Circulation 1997;95:1677-1682.

c. Supraventricular tachycardia (SVT) is specifically included in the non-shockable rhythm class, in accordance with AHA recommendations

b

and the AAMI standard DF80.

rhythm class

ECG test

sample

a

size

meets AHA recommendations

b

for adult defibrillation

observed 90%

one-sided

performance

lower

confidence limit

category

specifications

pads for defibrillation,

pacing, monitoring,

cardioversion

Disposable, adhesive pads with a nominal active surface area of 80 cm

2

each,

provided in a disposable plastic case, and an integrated 48 inch (121.9 cm),
typical, cable. Pads in case are designed to fit into carry cases.

SMART Pads II compatibility

defibrillator model

adult

patient use

infant/child patient use

FRx*

FR2/FR2+

FR/ForeRunner

MRx/XL/XLT/4000
HS1/OnSite/Home

competitive adapters

yes
yes
yes
yes

no; use M5071A

yes

yes

no, use M3870A

no

manual mode only

no; use M5072A

manual mode only

* Pre-connectable to FRx defibrillator only.

pads shelf life

Pads package is labeled with a use-by date of at least two years from date of
manufacture.