Philips 861304 User Manual

Page 53

Advertising
background image

45

Ph

ilip

s M

e

di

ca

l Sy

st

e

m

s

AP
P

E

ND

IC

ES

CPR Coaching

adult ventilation instruction

yes, no

yes

Optional CPR Coaching
includes rescue breaths at the
rate determined by the CPR
Coaching compression:
ventilation ratio for adults
when an adult pads set is
installed.

Note: if this parameter is
configured to NO, CPR Coaching
will always be compressions-only
when an adult pads set is
installed.

CPR Coaching

infant/child ventilation

instruction

yes, no

yes

Optional CPR Coaching
includes rescue breaths at the
rate determined by the CPR
Coaching compression:
ventilation ratio for infants and
children when an infant/child
pads set is installed.

Note: if this parameter is
configured to NO, CPR Coaching
will always be compressions-only
when an infant/child pads set is
installed.

CPR Coaching

compression:ventilation

ratio

• 30:2 adult and

30:2 infant/child

• 30:2 adult and

15:2 infant/child

• 15:2 adult and

15:2 infant/child

30:2 adult and
30:2 infant/
child

When the user presses the i-
button for optional CPR
Coaching during a protocol
pause or NSA pause, the FRx
will provide coaching in basic
CPR for cycles of 30
compressions and 2
ventilations for adults,
children, and infants. Pauses
begin and end with
compressions.

parameter

settings

default

default description

Advertising