Removing cooling filter, Care and maintenance – Philips V200 User Manual

Page 130

Advertising
background image

Chapter 10

10-8

Respironics V200 Ventilator Operator’s Manual

REF 1057983 A

Care and Maintenance

Table 10-1: Schedule for Periodic Maintenance

Removing Cooling Filter

Figure 10-4: Removing Cooling Filter

Refer to Figure 10-4.

1. Remove black filter retaining bracket (1). (Pry off if necessary.)

2. Wash filter (2) in soapy water.

3. Dry thoroughly before reinstalling.

12,500 hours

• 12,500 hour preventive

maintenance kit (P/N
1001733). Kit contents are
subject to change.

• Install 12,500 hour preventive

maintenance kit

• Clean ventilator interior and exterior
• Complete performance verification

procedure

CAUTION: The 12,500 hour preventive maintenance procedure is to be
performed only by a qualified service technician.

As required

• External oxygen sensor

• Backup Battery

• Replace and recalibrate new sensor by

running Extended Self-Test

• For charging and maintenance

instructions, see “Backup Battery” on
page 4-4
.

CAUTION:

Do not remove any screws from the cooling filter area. Removing screws
from this area will result in damage to internal components.