Chapter 13

13-104

Respironics V200 Ventilator Operator’s Manual

REF 1057983 A

Options and Accessories

•

Do not use with neonatal tracheostomy tubes.

•

Do not use on neonates.

•

Do not inflate the tracheostomy tube cuff until Speaking Mode has
been turned off.

•

Do not use on patients who are dependent on PEEP therapy.

•

Do not use with HME filters.

•

Discontinue using Speaking Mode promptly and institute appropriate
ventilation if the patient experiences difficulty or their status
deteriorates from baseline parameters.

•

Perform an airway patency test immediately after activating Speaking
Mode.

NOTE:

Speaking Mode is available ONLY in invasive ventilation mode.

Patient Preparation

Using Speaking Mode on a patient requires preparation before activating the
mode. Refer to the Sequence for Use diagram in Figure 13-58 and follow the
necessary steps before starting Speaking Mode and discontinuing Speaking
Mode. The Speaking Mode Clinical Guide describes patient care steps in
detail.

WARNING:

Read and understand these Speaking Mode Option instructions before using
Speaking Mode on a patient.

WARNING:

Wait at least 72 hours after a tracheostomy before initiating Speaking
Mode.