Respironics V200 Ventilator Operator’s Manual

11-1

Chapter 11. Diagnostics

The Diagnostic Mode allows you to:

•

Run short self test (SST)

•

Run extended self test (EST)

•

Run hardware diagnostics to help you troubleshoot SST or EST failures

•

Check the software revision of the ventilator

•

Set user configuration, including:
•

time and date

•

compliance compensation enable/disable

•

local altitude and time format

•

backup battery (confirm at startup) enable/disable

•

Check diagnostic codes.

WARNING:

The patient must be disconnected from the ventilator before entering the
Diagnostic Mode since normal ventilation is suspended.

CAUTION:

Troubleshooting and repair should be performed only by a qualified
service technician.

NOTE:

The “Hardware” function and EST in the Diagnostics Mode should only
be run by qualified personnel.