Rs-232 communications protocol – Philips V200 User Manual

Page 288

Advertising
background image

Appendix A

Appendix A-2

Respironics V200 Ventilator Operator’s Manual

REF 1057983 A

RS-232 Communications Protocol

Ventilation Report

Description

Example

Resolution

Range

Units

Comments

Command Name

VRPT

N/A

N/A

N/A

Number of characters
between the start and
stop codes

990

N/A

N/A

N/A

3 character field

Number of fields
between the start and
stop codes

134

N/A

N/A

N/A

3 character field

Start Code

0x02

N/A

N/A

N/A

ASCII Start
Transmission
Character (STX)

Time of request

13:45

N/A

N/A

N/A

24 hour clock,
hh:mm

Date

FEB

◆23◆1997◆ N/A

N/A

N/A

12 character field,
MMM

◆DD◆YYYY◆

Current Ventilation
Type

VCV

◆◆◆

N/A

VCV

◆◆◆

PCV

◆◆◆

NPPV

◆◆

N/A

The ventilation
type currently
being used by the
ventilator.

VCV Mode Setting

A/C

◆◆◆

N/A

A/C

◆◆◆

SIMV

◆◆

CPAP

◆◆

N/A

VCV Waveform Setting

RAMP

◆◆

N/A

RAMP

◆◆

SQUARE

N/A

VCV Patient Type

ADULT

N/A

ADULT

PED

◆◆◆

NEO

◆◆◆

VCV Respiratory Rate
Setting

12

◆◆◆◆

1

1 - 80

BPM

VCV Tidal Volume
Setting

500

◆◆◆

1

50 - 2500

mL

VCV Peak Flow Setting 30

◆◆◆◆

1

3 - 140

LPM

VCV PEEP Setting

0

◆◆◆◆◆

1

0 - 35

cmH

2

O

(hPa)

VCV Pressure Support
Setting

0

◆◆◆◆◆

1

0 - 100

cmH

2

O

(hPa)

VCV I-Trigger Type

PRESS

N/A

PRESS

FLOW

◆◆

AUTO

◆◆

N/A

VCV Pressure I-Trigger
Setting

2.0

◆◆◆

0.1

0.1 - 20.0

cmH

2

O

(hPa)

VCV Flow I-Trigger
Setting

3.0

◆◆◆

0.1

0.5 - 20.0

LPM

Table A-1: Ventilation Report (Sheet 1 of 10)