Glossary of symbols/controls

symbol

description

Pads cartridge handle. Green. Pulling the handle turns on
the defibrillator and opens pads cartridge for use.

Refer to operating instructions.

On/Off button. Green. Pressing the On/Off button when
the defibrillator is in standby mode turns the defibrillator
on; pressing and holding the On/Off button for one
second when the defibrillator is on turns the defibrillator
off and disarms the defibrillator. In addition, pressing the
On/Off button stops the battery insertion self-test that
automatically runs when a battery is inserted.

Information button (i-button). Blue. Pressing the i-button while it is
flashing during a patient care pause provides CPR Coaching; pressing it
while it is flashing and the defibrillator is chirping provides
troubleshooting guidance. Pressing it until it beeps at other times
provides summary information about the defibrillator’s last clinical use
and device status.

Caution light. Flashes during rhythm analysis, and is on
but not flashing when a shock is advised, as a reminder
not to touch the patient.

Shock button. Orange. Flashes when the defibrillator is
charged. If a shock is needed, the defibrillator directs the
user to press the Shock button to deliver a shock to the
patient.

Defibrillation protection. Defibrillation protected, type
BF patient connection.

Meets the requirements of the European medical device directives 93/
42/EEC.

Indicates that this device is optimized for Guidelines 2005.