HEARTSTART M5066A

Troubleshooting while the HeartStart is not in use

(green Ready light is not on)

behavior

possible cause

recommended action

chirps or

i-button flashes

• The battery power is low or the

SMART Pads cartridge needs to be
replaced.

• The defibrillator may have been

turned off without a pads cartridge
installed, or the installed pads
cartridge may not have its hard
cover in place.

• The training pads cartridge has been

left in the defibrillator.

• The defibrillator has been stored

outside the recommended
temperature range.

• The defibrillator has detected an

error during a self-test or cannot
perform a self-test, or the Shock
button is damaged.

• Press the blue i-button. Replace the

battery or pads cartridge if
instructed.

• Make sure the pads cartridge is

properly installed with the hard
cover in place. (See Chapter 5,
“Maintaining the HeartStart,” for
directions on installing the pads
cartridge.)

• Remove the training pads cartridge

and replace it with an Adult or
Infant/Child Pads Cartridge.

• Remove the battery for five

seconds then reinstall it to start
the battery insertion self-test. If it
fails, insert a new battery to repeat
the test. If it fails again, do not use
the defibrillator. If it passes, store
the defibrillator within the
recommended temperature range.

• Contact Philips for service if

needed.

no chirping and/or

i-button does not flash

• The battery is missing or completely

depleted.

• The defibrillator may have been

physically damaged.

Remove the battery for five seconds
then reinstall it to start the battery
insertion self-test. If it fails, insert a
new battery and repeat the test. If it
fails again, do not use the defibrillator.
Contact Philips for service.