HEARTSTART M5066A

energy*

(pediatric doses indicated

are based on CDC

growth charts for the

50th percentile weights

for boys.)

Using HeartStart Adult SMART Pads: 150 J nominal (

±

15%) into a 50 ohm load.

Using HeartStart Infant/Child SMART Pads: 50 J nominal (

±

15%) into a 50 ohm

load. Sample pediatric energy doses:

age

energy dose

newborn

14 J/kg

1 year

5 J/kg

2

− 3 years

4 J/kg

4

− 5 years

3 J/kg

6

− 8 years

2 J/kg

*

National Center for Health Statistics in collaboration with the National Center for Chronic
Disease Prevention and Health Promotion. CDC growth charts: weight-for-age percentiles, revised
and corrected
November 28, 2000. Atlanta, GA: Centers for Disease Control and Prevention ©
2000.

charge control

Controlled by Patient Analysis System for automated operation.

“charge complete”

indicator

Shock button flashes, audio tone sounds.

shock-to-shock cycle time

<20 seconds, typical, including analysis.

patient care

pause-to-shock time

Quick Shock. 8 seconds, typical, from end of patient care pause to shock
delivery.

disarm (AED mode)

Once charged, the defibrillator will disarm if:
• the patient’s heart rhythm changes to non-shockable rhythm,
• a shock is not delivered within 30 seconds after the defibrillator has charged

for shock delivery,

• the On/Off button is pressed and held down for at least one (1) second to

turn off the defibrillator,

• the adhesive pads are removed from the patient or the pads cartridge is

disconnected from the defibrillator,

• the battery is removed or is completely depleted or
• the impedance between pads is out of range.

adult shock

delivery vector

Via adhesive pads placed in the anterior-anterior (Lead II) position.

infant/child shock

delivery vector

Via adhesive pads typically placed in the anterior-posterior position.

category

specifications