Defibrillation Waveform

category

specifications

waveform parameters

Biphasic truncated exponential. Waveform parameters are automatically
adjusted as a function of patient defibrillation impedance. In the diagram at left,
D is the duration of phase 1 and E is the duration of phase 2 of the waveform, F
is the interphase delay (500 µs), and Ip is the peak current.
The HeartStart delivers shocks to load impedances from 25 to 180 ohms. The
duration of each phase of the waveform is dynamically adjusted based on
delivered charge, in order to compensate for patient impedance variations, as
shown below:

adult defibrillation

load

phase 1

phase 2

peak

delivered

resistance (

Ω) duration (ms) duration (ms) current (A)

energy (J)

25

2.8

2.8

65

128

50

4.5

4.5

40

150

75

6.25

5.0

30

155

100

8.0

5.3

24

157

125

9.65

6.4

21

159

150

11.5

7.7

18

160

175

12.0

8.0

16

158

pediatric defibrillation

(using M5072A infant/child reduced-energy defibrillator pads)

load

phase 1

phase 2

peak

delivered

resistance (

Ω) duration (ms) duration (ms) current (A)

energy (J)

25

4.1

2.8

28

35

50

5.1

3.4

20

46

75

6.2

4.1

15

52

100

7.2

4.8

12

54

125

8.3

5.5

10

56

150

9.0

6.0

9

57

175

9.0

6.0

8

55