of Thermocouples

Figures 9, 10, 11, 12, and 13 show the typical and maximum errors for the
different thermocouple types when used with the USB-9211/9211A over
the full temperature range. The figures also show the maximum error for
the thermocouple types with the USB-9211/9211A at room temperature,
15 to 35 °C. The figures account for gain errors, offset errors, differential
and integral nonlinearity, quantization errors, noise errors, and isothermal
errors. The figures do not account for the accuracy of the thermocouple
itself.

Temperature gradients across the USB-9211/9211A terminals affect the
cold-junction temperature accuracy. Refer to the

Cold-Junction

Temperature Measurement Accuracy

section for more information about

temperature gradients.