Figure 12. type k errors, Figure 13. type r and s errors, Cold-junction temperature measurement accuracy – National Instruments 9211 User Manual

Page 13: 12, and 13 sh, R to the, Cold-junction, Temperature measurement accuracy, Sectio

Advertising
background image

© National Instruments Corporation

13

USB-9211/9211A User Guide and Specifications

Figure 12. Type K Errors

Figure 13. Type R and S Errors

Cold-Junction Temperature Measurement Accuracy

Heat from other nearby heat sources can cause errors in thermocouple
measurements by heating up the terminals so that they are at a different
temperature than the cold-junction compensation sensor used to measure
the cold junction. The thermal gradient generated across the terminals can
cause the terminals of different channels to be at different temperatures, so
the resulting measurement creates errors not only in absolute accuracy but
also in the relative accuracy between channels.