Eneb (7), Eoip (10), Solutions – National Instruments NI-488.2 User Manual

Page 110: Windows, Mac os x, Linux

Advertising
background image

Appendix C

Error Codes and Solutions

NI-488.2 User Manual

C-6

ni.com

ENEB (7)

ENEB occurs when a GPIB interface is configured for use by the system,
but the driver cannot find the interface. This problem happens when the
interface is not physically plugged into the system, the I/O address
specified during configuration does not match the actual interface setting,
or there is a system conflict with the base I/O address.

Solutions

Make sure there is a GPIB interface in your computer that is properly
configured both in hardware and software using a valid base I/O address
by running the NI-488.2 Troubleshooting Utility, as follows:

Windows

1.

Select Start»Programs»National Instruments»
Measurement & Automation
.

2.

Select Help»Troubleshooting»NI-488.2 Troubleshooting Utility.

Mac OS X

Run Applications»National Instruments»NI-488.2»Troubleshoot.

Linux

Enter the following command:

<InstallDir>/natinst/ni4882/bin/gpibtsw

where

<InstallDir>

is the directory where you chose to install the

NI-488.2 software. The default is

/usr/local

.

EOIP (10)

EOIP occurs when an asynchronous I/O operation has not finished before
some other call is made. During asynchronous I/O, you can only use

ibstop

,

ibnotify

,

ibwait

, and

ibonl

or perform other non-GPIB

operations. If any other call is attempted, EOIP is returned.