National Instruments NI 9234 User Manual

Page 23

Advertising
background image

© National Instruments Corp.

23

NI 9234 Operating Instructions and Specifications

Accuracy

1

1

Refer to the

NI 9234 Circuitry

section for information regarding grounded signal

sources and measurement accuracy.

Measurement Conditions

Percent of

Reading

(Gain Error)

Percent of

Range

*

(Offset Error)

Calibrated max (–40 to 70 °C)

0.34%,

±0.03 dB

±0.14%,

7.1 mV

Calibrated typ (25 °C ±5 °C)

0.05%,

±0.005 dB

±0.006%,

0.3 mV

Uncalibrated max (–40 to 70 °C)

1.9%,

±0.16 dB

±0.27%,

13.9 mV

Uncalibrated typ (25 °C ±5 °C)

0.48%,

±0.04 dB

±0.04%,

2.3 mV

*

Range = 5.1 V

pk

Advertising