Physical, Maximum working voltage, Environmental – National Instruments NI 783xR User Manual

Page 53: Operating environment

Advertising
background image

Appendix A

Specifications

NI 783xR User Manual

A-8

ni.com

Physical

Dimensions (not including connectors)

NI PXI-783xR..................................16 cm by 10 cm (6.3 in. by 3.9 in.)

NI PCI-783xR..................................17 cm by 11 cm (6.7 in. by 4.3 in.)

I/O connectors.........................................Three 68-pin female high-density

VHDCI type

Maximum Working Voltage

Maximum working voltage refers to the signal voltage plus the
common-mode voltage.

Channel-to-earth

.....................................±12 V, Measurement Category I

Channel-to-channel

.................................±24 V, Measurement Category I

Caution

Do not use the NI 783xR for connection to signals in Measurement Categories II,

III, or IV.

Environmental

The NI 783xR is intended for indoor use only.

Operating Environment

NI 7830R, NI 7831R

40 MHz or 80 MHz timebase ..........0

°C to 55 °C, tested in

accordance with IEC-60068-2-1
and IEC-60068-2-2

NI 7833R

40 MHz timebase.............................0

°C to 55 °C, tested in

accordance with IEC-60068-2-1
and IEC-60068-2-2

80 MHz timebase.............................0

°C to 55 °C except the

following: 0

°C to 45 °C when

installed in an NI PXI-1000/B or
NI PXI-101X; tested in
accordance with IEC-60068-2-1
and IEC-60068-2-2

Advertising