Appendix A

Specifications

© National Instruments Corporation

A-9

NI 783xR User Manual

Relative humidity range ......................... 10% to 90%, noncondensing,

tested in accordance with
IEC-60068-2-56

Altitude................................................... 2,000 m at 25

°C ambient

temperature

Storage Environment

Ambient temperature range.................... –20

°C to 70 °C tested in

accordance with IEC-60068-2-1
and IEC-60068-2-2

Relative humidity range ......................... 5% to 95%, noncondensing,

tested in accordance with
IEC-60068-2-56

Note

Clean the device with a soft, non-metallic brush. Make sure that the device is

completely dry and free from contaminants before returning it to service.

Shock and Vibration (for NI PXI-783xR Only)

Operational Shock.................................. 30 g peak, half-sine, 11 ms pulse

Tested in accordance with
IEC-60068-2-27. Test profile
developed in accordance with
MIL-PRF-28800F.

Random Vibration

Operating ........................................ 5 Hz to 500 Hz, 0.3 g

rms

Nonoperating .................................. 5 Hz to 500 Hz, 2.4 g

rms

Tested in accordance with
IEC-60068-2-64. Nonoperating
test profile exceeds the
requirements of
MIL-PRF-28800F, Class 3.